This flat heating plate is improved and developed by referencing imported thermostatic equipment. The heating panel is integrally die-cast from aluminum alloy and coated with DuPont Teflon on the exterior. It delivers outstanding thermal conductivity with excellent thermal shock resistance and corrosion resistance. Its unique digital circuit design enables precise temperature selection & control, while clearly displaying both set temperature and real-time constant temperature value.
Flat Cast Aluminum Constant Temperature Heating Plate
1. Product Working Principle
The aluminum electric heating platform adopts tubular heating elements, bent into standardized shapes and cast into large flat plates with aluminum alloy molds. After precision machining, the cast aluminum heater surface is smooth and free of casting defects, enabling tight contact with heated workpieces. The surface load can reach 1–3 W/cm² with an operating temperature range from ambient up to 300℃, serving as a high-efficiency heater with uniform heat distribution. Premium aluminum alloy with superior thermal conductivity guarantees even surface temperature and minimal temperature difference across the plate. It features long service life, excellent thermal insulation, robust mechanical performance and anti-magnetic properties. Adding insulation covers on radiating surfaces can effectively reduce power consumption.
2. Key Product Features
- Heater formed with special molded aluminum alloy for flat, uniform temperature rise.
- LED digital temperature display on working surface with accurate readings.
- Fast, even heating speed, simple operation and safe to use.
- Typically used for heating & drying experiments of slides, flat substrates and samples on glass slides.
- Teflon-coated heating plate with superior heating performance.
3. Technical Specifications
- Operating environment: Ambient temperature -20℃ ~ +60℃, relative humidity < 80%
- Dielectric strength: Withstand 1500V 50Hz AC sine voltage test without breakdown within 1 minute
- Leakage current: < 0.5 mA
- Insulation resistance: > 2 MΩ
- Ground resistance: < 0.1 Ω
- Power tolerance: +5% ~ -10%
4. Model Size, Power & Weight List
| Model | Plate Size (mm) | Max Power | Net Weight |
|---|---|---|---|
| ET-1010 | 100×100×20 | 400W | 3.5Kg |
| ET-1515 | 150×150×20 | 600W | 5Kg |
| ET-2020 | 200×200×20 | 800W | 5Kg |
| ET-3020 | 300×200×20 | 1200W | 7Kg |
| ET-3030 | 300×300×20 | 1800W | 9Kg |
| ET-3625 | 360×250×20 | 1500W | 9Kg |
| ET-4020 | 400×200×20 | 1500W | 8Kg |
| ET-4030 | 400×300×20 | 1800W | 11Kg |
| ET-4040 | 400×400×20 | 2200W | 13Kg |
| ET-5050 | 500×500×20 | 3500W | 20Kg |
| ET-6020 | 600×200×20 | 2200W | 11Kg |
| ET-6030 | 600×300×20 | 2500W | 16Kg |
| ET-6040 | 600×400×20 | 3200W | 22Kg |
| ET-6050 | 600×500×20 | 3800W | 25Kg |

5. Application Scope
- Digital thermostatic stage & heating plate widely used for sample baking, drying and various thermal testing; essential equipment for biological, genetic, biochemical, analytical and university research laboratories.
- Also extensively matched on extrusion machines, injection molding machines, vulcanizing presses, hydraulic machines, ultrasonic cleaners, semiconductor production lines, cable manufacturing equipment, petroleum industry machinery and other industrial devices.
6. Operation & Maintenance Procedures
- Wipe the working surface completely before use; no water droplets, dust or foreign residues are allowed on the plate.
- Place sample flasks or lab containers on the heating plate.
- Connect power and turn on main switch. Press SET key when digits flash to enter setting mode; press ▲ to increase value, ▼ to decrease value, hold for fast adjustment. Press SET again to save and exit setting mode.
- A dedicated operator must supervise the equipment during heating operation.
- After finishing work, turn off heating switch and cut off power supply completely.
7. Safety Precautions
- Operating voltage shall not exceed 110% of rated voltage.
- All wiring terminals must be placed outside heating & insulation layers; equipment housing must be reliably grounded.
- Store cast aluminum heaters in dry environment. If insulation resistance drops below 1 MΩ after long storage, bake in an oven at ~200℃ for 5–6 hours to recover insulation performance.
- Secure fixed installation; effective heating surface must fully contact target workpiece. Dry running without load is strictly prohibited. Clean dust and contaminants on surface timely to avoid poor heat dissipation and shortened service life.
- Magnesium oxide powder inside heating element terminals shall be protected from moisture and pollutants to prevent electric leakage hazards.

8. Standard Heating Plate Operating Steps
- Connect power supply and turn on power switch; steady green power indicator confirms mains input.
- Turn on heating switch to start operation. Solid red heating lamp = continuous temperature rise; flashing red lamp = stable constant temperature mode with intermittent heating cycles.
- After use, turn off heating switch first. Green power light remains on as a hot surface warning – do not touch plate surface. Cut off main power only after the plate fully cools down (minimum 1 hour recommended).
9. Temperature Controller Display & Indicator Description
Basic Display State: After power-on, upper window displays PV measured temperature value, lower window displays SV target set value. If input measurement signal exceeds range (e.g. thermocouple open circuit), upper window alternately flashes “orA” and upper/lower limit values, and the controller automatically stops heating output.
Four LED indicator lights on front panel: OP1 (main output), AU1 (alarm 1), AU2 (alarm 2), RUN (running status).
10. Custom Heating Plate Solutions
Custom flat heating platforms of all sizes and materials are available: stainless steel, cast aluminum, cast copper, cast iron, 6061 aluminum alloy. Maximum standard platform size reaches 2m × 1.3m, all custom dimensions supported on demand.