PID digital display for precise temperature control, surface temperature up to 350℃.
Equipped with digital speed display, dual-screen display, and auto-tuning function.
Features high measurement accuracy, low temperature overshoot, one-touch operation,
internal/external thermocouple temperature measurement, thyristor control output, and thermocouple break protection.
I. Features of Digital Magnetic Stirrer
- Exclusive square TU2 oxygen-free copper plate with vacuum coating / black crystal ceramic heating panel, elegant and corrosion-resistant.
- Unique patented heating method, maximum surface temperature up to 350℃.
- Digital display for rotation speed.
- Fuzzy PID control algorithm, dual-screen digital display, auto-tuning, high precision, low temperature overshoot, one-touch operation, internal & external thermocouple, thyristor output, 160-240V wide voltage, and thermocouple break protection.
- Suitable for heating and stirring 50-10000ml standard or non-standard reaction flasks.
- Adopts German PAPST series DC brushless motor, stable performance, low noise, long service life, no sparks.
- Flame-retardant reinforced PBT injection molded housing, high temperature resistance, corrosion resistance, and good insulation.
- 30° inclined control panel for comfortable sitting or standing operation.
- Stepless speed regulation, stable at low speed, powerful at high speed.

ET-JB Intelligent Digital Magnetic Stirrer Specifications
| Power Supply | ~220V, 50Hz |
| Speed Range | 50-1800 rpm, stepless & digital display |
| Motor Power | 40W, DC14-24V |
| Heating Power | 260W |
| Max Surface Temp | 350℃ |
| Temp Control | Intelligent digital, thyristor control |
| Temp Accuracy | ±1℃ |
| Heating Wire | Cr20Ni80 |
| Heating Area | 140*140mm |
| Material | TU2 Oxygen-free Copper / Black Crystal Ceramic |
| Stirring Volume | 50-500ml beaker |
| Working Time | Continuous |
| Dimensions | 265*155*120mm |
| Package Size | 315*205*170mm |
| Weight | 2200±200g |
II. Safety & Operating Notes
- DO NOT operate without liquid (dry heating).
- For safe use, please ground the device.
- The motor is fan-cooled. During high-temperature heating, the motor MUST be running (at low speed/idle) to prevent overheating damage. Damage caused by improper operation is not covered by warranty.
- After high-temperature heating, turn off heating first, wait for cooling, then turn off stirring.
- Heating surface is hot during operation; avoid contact to prevent burns.
- Dry the unit completely if wet, spilled, or used in high humidity to avoid electric shock.
- Store in dry, non-corrosive environment when not in use for long periods.
- Ground the device if induced current occurs in high humidity; ensure ventilation.
- Ambient humidity: 35%-85% (non-condensing).
- Fuse: Φ5×20 15A.
Specifications are subject to change without notice due to product improvements.
Operating Instructions
- Fix the support rod, clamp the flask, insert the stirring bar and sensor plug, connect to 220V power, and turn on the main switch.
- Temperature Setting: Press the heating switch. Press SET, SV value flashes. Adjust with < > ∨ ∧ keys. Press SET again to confirm. Green light = heating; off = stop.
- Speed Setting: Press the stir switch. Press SET to adjust r/min value, confirm with SET.
- Auto-Tuning (AT): Minimizes temperature overshoot and improves stability. Re-set after changing solution or conditions.
- Start AT: Hold < button for 8 seconds. AT light flashes. Hold again for 8s to stop.
- The unit exits setting mode automatically if no operation for 8 seconds.
- A rubber cap protects the external thermocouple socket. Remove it when using external sensor.
- Thermocouple break protection: If “”hhhh”” appears and green light off, check the thermocouple connection.