High Precision Digital Constant Temperature Cooling Stage
Table of Contents
1. Preface
Thank you for purchasing our ET-360 intelligent digital constant temperature cooling stage. Please read this manual completely before powering on and installing the equipment. Strictly follow operational specifications to prevent equipment damage and safety hazards caused by improper operation. Keep this manual properly for future commissioning and maintenance reference.
If you encounter inconsistent functions or abnormal faults during operation, please contact our after-sales team or authorized distributors promptly.
⚠ Important Notice:The manufacturer reserves the right to update this manual and adjust technical parameters without prior notification.
2. Product Overview & Features
2.1 Product Overview
The ET-360 cooling stage is a split-type semiconductor thermostatic device, composed of an aluminum alloy cooling work platform and an independent intelligent temperature controller. An NTC temperature sensor collects the surface temperature of the platform in real time. The controller automatically analyzes temperature data and drives the TEC semiconductor module to switch between cooling and heating for bidirectional constant temperature control. It is widely used in laboratory constant temperature testing, low-temperature rework of electronic components, optical sample thermostatic experiments and other applications.
Figure 1: Actual photo of ET-360 cooling stage
2.2 Eight Core Advantages
- Automatic bidirectional switching between cooling and heating; the cooling/heating module starts and stops automatically without manual mode switching
- Wide temperature control range with high control precision to meet thermostatic requirements across multiple scenarios
- LED digital display showing real-time temperature, parameter codes and fault alarms
- Wide input voltage compatibility with strong power supply adaptability
- Independent fan control; delayed heat dissipation after shutdown to protect the semiconductor module
- Built-in standard RS485 communication interface, supporting remote temperature control and batch temperature data collection
- Audible and visual alarms for sensor failure, over-temperature and under-temperature conditions
- Stainless steel housing protection; cage-type screw terminals ensure convenient and firm wiring to extend service life
4. Controller Panel Description
Figure 2: Schematic diagram of temperature controller panel markings
① Temperature Sensor Probe: Collects real-time surface temperature of the cooling platform
② LED Display: Shows real-time temperature, parameter codes and fault codes HHH (over-temperature) / LLL (under-temperature)
③ Buzzer: Triggers alarm sound for equipment faults, parameter setting errors, over-temperature and under-temperature events
④ Four Status Indicator Lights:
S/TA: Setting status & temperature alarm indicator
C/FD: Cooling running indicator; flashes during fan delay period
H/FD: Heating running indicator; flashes during fan delay period
RS485: Communication status indicator
⑤ Operation Buttons:
RET: Return to temperature display screen / Power on & off
SET: Enter parameter menu / Confirm parameter modification
UP: Increase parameter code / numerical value
DOWN: Decrease parameter code / numerical value
⑥ Wiring Terminals: POWER power supply, TEC semiconductor module, FAN fan, RS485 communication, NTC sensor, S1/S2 external over-temperature protection terminals
5. Basic Operation Instructions
5.1 Power ON / OFF of the Temperature Controller
The controller powers on automatically after energization. Press and hold the RET key for more than 2 seconds to power off the unit. Short press RET once to power on when the unit is shut down. The RET key will not respond within 5 seconds immediately after power-off.
5.2 Parameter Table
| Symbol | Function Description | Setting Range | Factory Default | Unit |
|---|---|---|---|---|
| E1 | Cooling Start Temperature | -40 ~ +120 | 19 | ℃ |
| E2 | Heating Start Temperature | -40 ~ +120 | 17 | ℃ |
| E3 | Cooling Hysteresis | 0.1 ~ 20 | 1 | ℃ |
| E4 | Heating Hysteresis | 0.1 ~ 20 | 1 | ℃ |
| E5 | Temperature Alarm Deviation | 10 ~ 50 | 50 | ℃ |
| E6 | Fan Delay Stop Time | 0 ~ 60 | 5 | min |
| E7 | Temperature Calibration Value | -10 ~ +10 | 0 | ℃ |
| E8 | Communication ID | 1 ~ 128 | 1 | / |
5.3 Control Logic Example (Factory Default Parameters)
Cooling starts when temperature exceeds 19℃ (E1); cooling stops when temperature drops to 18℃ (E1-E3), and the fan continues running for 5 minutes before shutdown. High-temperature alarm triggers at 69℃ with HHH displayed on screen and buzzer activated.
Heating starts when temperature falls below 17℃ (E2); heating stops when temperature rises to 18℃ (E2+E4), and the fan keeps running for 5 minutes before shutdown. Low-temperature alarm triggers at -33℃ with LLL displayed on screen and buzzer activated.
6. Parameter Setting (E1~E9)
6.1 E1 Cooling Start Temperature
Under standby status, press and hold SET for 2 seconds to enter the menu. Use UP/DOWN to select E1, then press SET to enter value adjustment. Press and hold UP/DOWN for fast value change. Restriction: E1 ≥ E2+E4.
6.2 E2 Heating Start Temperature
Under standby status, press and hold SET for 2 seconds to enter the menu. Use UP/DOWN to select E2, then press SET to enter value adjustment. Press and hold UP/DOWN for fast value change. Restriction: E2 ≤ E1-E3.
6.3 E3 Cooling Hysteresis
Cooling output will turn off when measured temperature < E1-E3. Adjustment operation is the same as above.
6.4 E4 Heating Hysteresis
Heating output will turn off when measured temperature > E2+E4. Adjustment operation is the same as above.
6.5 E5 Temperature Alarm Deviation
Audible & visual alarm will activate and HHH/LLL will show on screen when temperature>E1+E5 or temperature<E2-E5.
6.6 E6 Fan Delay Stop Time
The fan continues heat dissipation after cooling/heating stops; the corresponding indicator flashes during the delay period.
6.7 E7 Temperature Calibration
Used to correct temperature measurement deviation. Calibrated temperature = measured temperature + calibration value; both positive and negative values are supported.
6.8 E8 Device Communication ID
RS485 communication identification number, adjustable from 1 to 128; factory default value is 1.
6.9 Restore Factory Defaults
Under standby state, press the RET key twice consecutively to restore all parameters to factory default values with one click.
7. Safety Precautions
- Cut off all power supply completely before wiring to avoid short circuit and burnout of the controller and TEC module.
- Route the NTC sensor cable away from high-power power cables to prevent electromagnetic interference leading to inaccurate temperature readings.
- The TEC semiconductor module has positive and negative polarities; reverse connection is strictly prohibited.
- Shielded twisted-pair cable is recommended for RS485 communication wiring.
- Do not operate the unit long-term in high humidity, corrosive gas or heavy dust environments.
- The external load must not exceed the rated power of the equipment.
- Non-professional personnel are forbidden to disassemble the housing or modify internal wiring.
- Disconnect the main power supply if the equipment will remain idle for a long period.
8. Fault Codes & Troubleshooting
- Display HHH: Over-temperature alarm → Check cooling fan and E5 alarm parameter setting.
- Display LLL: Under-temperature alarm → Inspect the TEC cooling module and ambient ventilation condition.
- Temperature reading drift: Check NTC probe wiring and adjust E7 temperature calibration value.
- No cooling or heating output: Verify E1/E2 temperature thresholds and confirm the controller is powered on.
- Fan fails to run: Check FAN terminal wiring and E6 fan delay parameter.
9. Packing List
• ET-360 Constant Temperature Cooling Stage Main Unit ×1
• Operation Manual ×1
10. Warranty Terms
The whole unit is covered by a 12-month warranty under normal operating conditions. Free repair and component replacement will be provided for non-human damage within the warranty period. Damage caused by accidental dropping, water ingress, unauthorized modification or overload operation is excluded from warranty coverage.